Search
  
Categories
Manufacturers
Top » Catalog » Laversab » » Model 6500

Model 6500: Automated Air Data Test Set

   
Specifications
Features
   

Static Output

Pressure function

  • range: 0.1 to 42 inHg
  • resolution: 0.001 inHg
  • accuracy: ?0.002 inHg
  • units: inHg, mbar

Altitude function

  • range: -4000 ft. to 100,000 ft.
  • resolution: 1 foot
  • accuracy: ? 2 ft. @ 0 ft.
  • ? 6 ft. @ 35,000 ft.
  • ? 12 ft. @ 50,000 ft.
  • ? 50 ft. @ 80,000 ft.
  • units: feet, meters

Climb function

  • range: 0 ft/min to ?50,000 ft/min
  • resolution: 1 ft/min
  • accuracy: ?1% of rate of climb
  • units: feet/min, meters/sec

Pitot Output

Pressure function

  • range: 0.1 to 110 inHg
  • resolution: 0.001 inHg
  • accuracy: ?0.003 inHg
  • Pt units: inHg, mbar, EPR
  • Qc units: inHg, mbar

Airspeed function

  • range: 0 to 1100 knots
  • resolution: 0.1 knots
  • accuracy: ? 1.5 knots @ 20 knots
  • ? 0.5 knots @ 50 knots
  • < 0.25 knots above 100 knots
  • < 0.1 knots above 300 knots
  • units: knots, km/hr, mph, Mach
  • Rate of change: 0 to 500 knots/min

Mach function

  • range: 0.0 to 5.0 Mach
  • resolution: 0.001 Mach
  • accuracy: ?0.001 above 0.2 Mach

Dimensions & weights

  • 19-inch rack-mount kit optional

Environmental specs

  • Operating temp. 0?C to 50?C
  • Storage temp. -25?C to 75?C
  • Humidity: 5 to 95% non-condensing
RVSM compliant:
The very high accuracy of the tester and NIST-traceable calibration make the tester fully RVSM-compliant.

Calibration period: 1 year

Automated operation
Simply enter the target set-points for altitude and airspeed and the unit automatically achieves the targets, ramping at the programmed rates. Control at the set-point is stable within ? 1 foot, ? 0.1 knot or ?0.001 inHg.

Programmable limits
User-programmable limits for altitude, airspeed, Mach and VSI provide protection to the aircraft.

“Jog” and “Hold” features
Set-points can be “jogged” up or down in small increments. Ramping to a set-point can be put on “hold” and controlled at the current measured value.

“Ground” feature
Brings Pitot and Static outputs to ambient “ground” pressure after completing a test.

?Profiles? feature
A profile of set-points can be downloaded from a PC, allowing the user to step through the set-points using a single key. Each profile can have up to 50 set-points and 20 profiles can be stored in the unit.

Leak checks
Pitot and Static leak-checks can be performed either independently or simultaneously. Three leak-timers can be programmed to perform timed leaks between 1 to 20 minutes. The display of ?instant? leak rates provides immediate feedback to the user.

Pressure & Vacuum requirement
External pressure (Dry Air only) and vacuum supplies are required. The PVS is ideally suited for providing these supplies

Local display and keypad
The local display and keypad are the operator interface for the tester. All valid parameters, including altitude, climb and airspeed, are clearly displayed simultaneously on a single screen on the local display.

Applications

  • Ideal tester for use in an instrument-shop for testing and calibrating ADC’s, altimeters, airspeed indicators, VSI’s, cabin pressure controllers, FDR’s and other air-data instruments.
  • Can be rack-mounted in ATE systems

Power requirement

  • 90-260 VAC, 47-440 Hz ; 100 VA

External Interface

  • RS232 and IEEE-488
  • Remote unit optional
 
Downloads:
 
Datasheet
Manual

.
.
This product was added to our catalog on Tuesday 12 September, 2017.
.
.